DAQmxSetup1502Cal
int32 DAQmxSetup1502Cal (CalHandle calHandle,
const char channelName[], float64 gain);
Purpose
Sets the SCXI-1502 module to the specified gain
value. Calibration I/O points can be measured by supplying
reference signals to the specified channel and measuring the
outputs. Specify each of these points using the DAQmxAdjust1502Cal function.
|
Note The terminal where the module output is
measured depends on the configuration of the module in MAX.
National Instruments recommends cabling the module to the digitizer
so that the output appears on the MCH0± pins of the rear signal
connector. Refer to the module user manual for more information on
the routing of module output. |
Parameters
Input
|
Name |
Type |
Description |
calHandle
|
CalHandle |
A reference to the calibration session that you
created using the DAQmxInitExtCal function. |
channelName
|
const char [] |
The physical channel to calibrate. |
gain
|
float64 |
The gain value to calibrate. |
Return Value
Name |
Type |
Description |
status
|
int32 |
The error code returned by the function in the
event of an error or warning. A value of 0 indicates success. A
positive value indicates a warning. A negative value indicates an
error. |
|